Blood Microsampling for Complete Blood Count: Take Heed of Preanalytical Errors
ISSN: 1929 - 4247Publisher: author   
Blood Microsampling for Complete Blood Count: Take Heed of Preanalytical Errors
Indexed in
Health Sciences
ARTICLE-FACTOR
1.3
Article Basics Score: 2
Article Transparency Score: 2
Article Operation Score: 2
Article Articles Score: 3
Article Accessibility Score: 2
SUBMIT PAPER ASK QUESTION
International Category Code (ICC):
ICC-0702
Publisher: Lifescience Global Inc.
International Journal Address (IAA):
IAA.ZONE/192991244247
eISSN
:
1929 - 4247
VALID
ISSN Validator
Abstract
Background: The credibility of the result of a complete blood count is closely connected with the preanalytical phase. Objectives: This study evaluated accordance of filling of microtubes with manufacturer’s recommendation and assessed the effect of storage of overfilled and underfilled samples on the results of complete blood count. Design and Methods: Volume of blood samples collected into microtubes in the wards of the University Children's Hospital in Cracow during one month was analyzed. In the stability studies, overfilled and underfilled samples stored at ambient temperature were analyzed at 1, 2, 3 and 12 hours after phlebotomy. The analysis was made using the SYSMEX XT-1800i analyzer. Results: More than half of the analyzed samples were incorrectly filled. 63% of the samples were filled above the manufacturer's recommended volume and 15% of test-tubes were filled below the recommendation. We observed differences between colle...